
Add to Cart
IC6200 Electrolytic Regeneration Suppression Technology Ion Chromatography Instruments
Introduction of Ion Chromatography Instrument
Ion Chromatography is a well established ion exchange chromatography technique for the analysis of anions and cations. In contrast to general HLPC systems, often suitable for a wide range of applications, ion chromatographs are chromatography systems specifically adapted for the analysis of anions and cations. Metal-free flow paths, conductivity detection and ion suppression of eluent ions are a few of the specific necessities to enable reliable chromatographic analysis of anions, cations, organic acids, transition metals and amines.
Features of Ion Chromatography Instrument
1. Intergrated apperance design, high brightness LCD full touch GUI controller
2. High degree of automation "One Key Rinse" and "One Key Maintenance" automatic maintenance function.
3. Higher intergration: Support optional parts upgrade to 2D system and multi-detector combination.
4. Full PEEK pump and electric injection valve.
5. Powerful and convenient chromatpgraphy workstation
6. High precision temperatur-controlled conductivity detector.
Parameters of Ion Chromatography Instrument
Wayeal IC6200 Integrated Ion Chromatography Instrument for Cations or Anions Analysis | |
System Parameter | |
Qualitative Repeatability | ≤0.5% |
Quantitative Repeatability | ≤1% |
Pump System | |
Flow Rate | (0.001~10.000)mL/min |
Pump Head Pressure Resistance | (0 ~ 35) MPa |
Flow Rate Setting Inaccuracy | <0.1% |
Flow Stability | <0.1% |
Pressure Display Accuracy | 0.01Mpa |
Pressure Pulsation | <0.5% |
Conductivity Detector | |
Baseline Noise | ≤0.1%FS |
Baseline Drift | ≤0.5%FS |
Minimum Detectable Concentration (Cl) | ≤0.0002μg/mL |
Minimum Detection Concentration (Li) | ≤0.0001μg/mL |
Conductivity Cell Volume | ≤ 0.6μL |
Conductivity Detection Range | 0-15000μS/cm |
Detector Resolution | 0.00238nS/cm |
Conductivity Cell Temperature Setting Error | ±0.01°C |
Resolution of Conductivity Cell Temperature Display | 0.001°C |
Suppressor | |
Constant Current Source Range | 0-500mA in 0.1mA increments |
Column Oven | |
Temperature Range | Room temperature +5°C~85°C |
Control Temperature Stability | ≤ 0.1°C |
Autosampler | |
Number of Sample Vials | ≥ 120 (1.5mL standard sample bottle) |
Repeatability | Full loop injection < 0.3% RSD, Partial injection < 0.5% RSD, Non-destructive injection < 1 % RSD (injection volume > 5uL) |
Syringe Size | Standard: 500μL; optional; 250μL, 1000μL, 2500μL |
Loop Size | Standard: 100μL; optional: 20μL, 50μL, 200μL |
Linearity | >0.999 |
Cross-contamination | <0.01% |
Eluent Generator | |
Flow Rate Range | 0.001 to 3.000ml/min |
Eluent Concentration Range | KOH, MSA-0.1 to 100mM |
Concentration Accuracy | 0.1mM |
Original Concentration of Electrolyte | 25% KOH/MSA |
Volume of Electrolyte | 1000mL |
Maximum Working Pressure | 21MPa (3000psi) |
Gradient Accuracy | 1.5% |
Industry Application
Application of ion chromatography in ambient air
1. Fixed source waste: detection of acid mist
Application of ion chromatography in soil and solid waste
1. Detection of anions in solid waste
2. Detection of cyanide and sulfide in solid waste.